Patient Monitor Manufacturer – AKAS TRAZE 12A

patientmonitor-t12a
Product Overview

Patient monitors found profound use in critical care units and other healthcare departments to constantly monitor body temperature, pulse rate, respiratory rate and blood pressure. The periodic check on these key parameters helps in live monitoring the state of the patient. We make sure patients don’t feel uncomfortable with lots of probes and wires running over them. The non-contact temperature feature in the device monitors patients once in 4 or 6 hours. The device records and provides the trend of various readings over a period of time.

Features
  • Slim and compact design
  • Audio visual alarm
  • High contrast display
  • CMS connectivity through Wi-Fi
  • Touch screen
  • Digital Enhancement Technology
  • Machine can totally be operated with *touch screen*, *keys* or *jog dial*. Even if one fails other 2 options will work
  • Slim model; does not take much space
  • Wall mountable – like a photo frame
Technical Specifications
  • Battery : 14.4V NiMah
  • Battery Capacity : 2 hours
  • Power Consumption : 21.5W
  • Input voltage : 220-240V Ac, 50/60HZ
  • Display : 12 ʻinchʼ TFT display with resistive touch

SpO2

  • Measuring range : 0-100%
  • Resolution : 1%

Pulse rate

  • Measuring range : 30-254BPM
  • Resolution : 1 BPM

Non-Invasive Blood Pressure (NIBP)

  • Method : Oscillometric
  • Mode : Auto,Manual,STAT
  • Measuring interval in Auto Mode : 1,2,4,……
  • Repeat cycle update in Stat Mode : 1-5
  • Systolic Range : Adult – 30 to 250 mmHg
    Pediatric – 30 to 250 mmHg
    Neonate – 20 to 160 mmHg
  • Diastolic Range : Adult – 10 to 180 mmHg
    Pediatric – 10 to 150 mmHg
    Neonatal – 10 to 140 mmHg
  • Mean Range : Adult – 10 to 180 mmHg
    Pediatric – 10 to 180 mmHg
    Neonatal – 20 to 200 mmHg
  • Over pressure protection : Adult mode (297+/-3)mmHg
  • Resolution : 1mmHg
  • Alarms : High & low exceed alarm for SpO2 & pulse rate, Sensor failure, Low perfusion,
    Battery low indication,High & Low exceed alarm for NIBP ( Sistolic, Diastolic, Mean)

ECG

  • Lead type : Standard 5-lead/3-LEAD
  • Lead selection : I, II, III ,aVR, aVL, aVF, V
  • HR measurement range : 30 bpm ‒ 300 bpm
  • Sweep speed : 12.5 mm/s, 25 mm/s, 50 mm/s
  • Gain selection : 2.5mm/mv,5,10,20mm/mv Auto
  • Bandwidth : Surgery mode: 1-20HZ
    Monitor mode – 0.5-40HZ
    Diagnostic mode – 0.05-75HZ

RESPIRATION

    • Method : Impedance between R-F(RA-LA)
    • Bandwidth : 0.3/2.5 HZ AC
    • Respiration rate : Adult: 0-120rpm

Neo/Ped – 1rpm

  • Resolution : 1rpm

 

Temperature

  • Channel : 1/2
  • Measuring range : 0-50deg C
  • Resolution : 0.1deg C

Fast Temperature

  • Channel : 1
  • Measuring range : 0-50deg C
  • Resolution : 0.1deg C

Accuracy

  • SpO2 Adult/Pediatric : 70-100% +/- 2%
    0-70% Unspecified
  • Neonate : 70-100% +/- 2%
    0-70% Unspecified
  • Pulse Rate : Adult/Pediatric 30-250 bpm +/- 2 bpm
    Neonate 30-250 bpm +/- 2 bpm
  • Pulse Rate : ± 1 at 40-200 bpm
  • NIBP : ± 3mmHg
  • ECG : ±0.02mV or ±5%
  • Respiration : +/-2rpm
  • Temperature : +/-0.1 deg C
  • Fast Temperature : +/-0.1 deg C
  • Dimension & weight : 294mm(L) x 97mm(H) x 267mm(W),
    3.4kg