| Model
|
Venus
- 3
|
| Display |
8.4” Color TFT Display
|
| Trace
|
7
Waveforms Maximum
|
| Patient Range |
Adult,
Pediatric and Neonatal |
| Power |
12V
DC 50/60 HZ,
80VA
|
| Battery
Backup |
3
Hrs with 12V/3.7Ah Li Battery
|
| Trend
time
|
1-72
Hours
|
| Alarm
|
audible and visual alarm |
|
Recorder
(optional)
|
Built
in, thermal array, 3 channels |
| Other |
Drug Calculation,
Oxy-CRG, 100 Alarm History
|
|
ECG Lead Type
|
5-
Lead
|
| Heart
Rate
Range |
15 -300 bpm
|
| Bandwidth |
Surgery
: 1 – 15 Hz
Monitor
: 0.5 – 35 Hz
Diagnostic : 0.05 –
100 Hz |
| S-T
Segment
Range |
-2.0mV
to +2.0 mV
|
| Arrhythmia analysis
|
13 Lethal Arrhythmias
detected & displayed
|
|
RespirationTechnique
|
Trans-Thoracic
Impedance (RA-LL)
|
| Measurement
Range |
0-150 rpm
|
| Apnea |
0-40 sec
|
|
NIBP Method
|
Oscillometric |
| Operation
Modes
|
Manual/Automatic/STAT
|
| Measurement Types
|
Systolic, Diastolic, Mean
|
| Measurement
Range |
Adult
:
10-270 mmHg
Infant
:
10-200
mmHg Neonatal : 10-135 mmHg |
|
SpO2
Range |
0-100% |
| Accuracy |
80 ~ 100%± 2%; 0 ~ 69%
unspecified |
| Pulse rate
|
Range: 0-250
bpm |
| Accuracy |
±1 bpm (30-230
bpm) |
|
Temperature Technique |
Resistance |
| Channel |
1 channel/
2channel |
| Scale |
0C
and 0F selectable |
| Measurement
Range |
00C - 50 0C
|
|
IBP
(optional) Measurement
Range |
-50 ~ 350
mmHg |
| Channel
|
2 channels
|
|
EtCO2
(Optional) Method |
Built
in sidestream sensor |
| CO2
Display
Range
|
0-99 mmHg
|
| CO2 Accuracy
|
+/- 3mmHg (0-40
mmHg)
+/- 8% of reading (41-76 mmHg)
+/-10% of
reading(77-99 mmHg |
| Flow
Rate
Range
|
User-selectable
50,100, 150
ml/min
|