AKAS Rhythm A4 Product Specification |
Recording Format |
12, 6, 3 channel |
Recording Mode |
Automatic and Manual mode |
Recording Speed |
25, 50 mm/s |
Recording Paper |
216mm x 30m, roll type, Grid and non-grid paper |
Printer |
High resolution Thermal Array Printer with 8 dots/mm Resolution |
Paper Sensor |
Out of paper indication |
Display |
Graphic LCD |
ECG Display |
Lead: standard 12 leads Sweep speed:25, 50 mm/s Sensitivity: 5, 10 mm/mV |
CMRR |
Greater than 100db |
Input Impendence |
Greater than 40M Ohm |
Frequency Responce |
Filtered: 0.5 - 30Hz(-3 dB) Unfiltered: 0.05 - 150Hz(-3dB) |
| Recording ECG Leads |
12 standard leads: I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
1 to 100 drop /min for Micro |
| Electrode Sensor |
Electrode disconnect indication |
ADC Sample Bit |
12bits |
ADC Sampling Rate |
500Hz |
| ECG Memory |
10 sec for each lead |
| Calculations |
Full calculation for all 12 leads |
Interpretation |
Yes |
Stress ECG interface |
Optional |
Automatic power OFF operation |
5min,form the last operation |
Type protection against electric shock charging period |
Class I(Degree of production against shock:Type CF applied part) |
Defibrillator protection type |
5000V, EN, IEC 60601-2-25 Requirements |
Patient Leakage current |
Less than 10 u A |
Power Requirement |
100-240VAC +/- 10%, ~50/60 Hz free voltage |
Power consumption |
Approx 45 VA |
Rechargeable Battery |
24V, 1200 mAh, NiCd, More than 1.5 hour continuous recording |
Battery Charger |
Battery Charger status is indicated by LED. Approximate 5 hours is required to charge fully |
Dimension(W x H x D) |
333x310x87(Unit:mm) |
Net weight |
3.4 kg |
External Output(optional) |
RS232C |
| |
|
|