Rhythm 350 

Read Specification
FEATURES
  • High-resolution thermal printing array system
  • Frequency responce is a high as 150Hz
  • Capable of printing continuously on channel trace and annotations including leadmark and parameter such as sensitivity, paper speed and filter operation status
  • High resolution digital filter inhibits baseline drift without affecting ECG waveform
  • One touch operation
  • Concept of floating input circuit meets IEC safety standard
  • Rechargeable battery supporting more than 50 patients
  • Lead standard 12 leads


SPECIFICATIONS
AKAS Rhythm 350 Product Specification

Work Environment 

Temperature

+5C ~ +35C 

Relative Humidity

Greater than 80%

Power Supply

Main Unit 220V+20V, 50Hz+1Hz
Under direct power supply the device can work for half an hour

Atmospheric pressure

86KPa~106KPa

Noise level

<15 Vp-p

Cmrr

>60dB

Range frequency character

10Hz reference mark, 1Hz~75Hz(-3dB,+0.4dB)

Low frequency character

Time constant >3.2s

Sensitivity

Standard sensitivity is 10mm/mV, +0.2mm/mV

Sensitivity control 5, 10, 20mm/mV, +5%
Paper speed 25mm/s, 50mm/s, +5%
Safety standard Class I, Type CF
Input Power Greater than 150VA

Fuse Specification

AC 220V/2A

 