| Model
|
SP
410
|
| Power supply |
AC
180V to 270V 50Hz |
| Power
Consumption
|
7
VA
|
| Battery |
4
hrs @ 5ml/Hr flow rate on 50 ml syringe with 12V NiCd
Battery
|
| Applicable Syringe
|
10ml,
20ml, 30ml, 50ml &100ml Syringe from JMS, Braun,
Dispovan, BD and Custom programmable
for any brand
Syringe
|
|
Flow rate Range
|
0.1
– 240 ml/Hr (10ml Syringe) 0.1 – 400 ml/Hr (20ml
Syringe) 0.1 – 600 ml/Hr (30ml Syringe) 0.1 – 840
ml/Hr (50ml Syringe) 0.1 – 999 ml/Hr (100ml
Syringe
|
| Infused Volume
|
0.1 – 999.9 ml
|
| Elapsed Time
|
Display of Elapsed
time from 00.00 to 99.59 hrs
|
|
Accuracy
|
Mechanical
Accuracy ±1% Including Syringe ± 3%
|
| Bolus |
Purge Bolus & Auto
Bolus
|
| Bolus Rate
|
200
ml/Hr 10 ml Syringe 360 ml/Hr 20 ml Syringe 560
ml/Hr 30 ml Syringe 800 ml/Hr 50 ml Syringe 999
ml/Hr 100 ml Syringe
|
| Bolus Volume
|
0.1ml to 5 ml in 0.1
ml increments
|
| Sequential Mode
|
Sequential program of
five different flow-rate at different time or volume
target
|
| KVO |
0.0ml/Hr to 9.9ml/Hr
in 0.1 ml increment
|
| Occlusion Pressure
|
10 steps selection
|
| Alarms
|
Syringe holder
distributed during infusion, Target Volume reached,
Target time reached, Motor unable to drive, End of
Syringe, PCA probe disconnected, Occlusion
|
| Modes |
Flow
Rate only Flow Rate & Target volume Flow
Rate & Target Time Target volume and target time
|
| Audio Visual
Indication
|
Near
Empty, Battery
low, Operation on battery |
| Special Functions
|
Adjustable buzzer
volume(OFF – 9 level)
|
| PCA(Optional) |
PCA dose can be
present from 0.1 to 10ml Minimum locking time interval
between two doses Time interval from 0 - 24hrs in steps
of 1 min Delivered and non delivered doses display
|
| Safety Testing
|
IEC 601-01
|
| Dimension
|
275 x 161 x 112
|
| Weight
|
2.5kg
|
| Drug Library
|
Yes |
| Display
|
LED
& LCD
Display
|